You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Hello, I am testing keyoscacquire with a Keysight DSOX1202G and when issuing the scope = koa.Oscilloscope('LONG ADDRESS') command, the library responds with: Connected to: KEYSIGHT TECHNOLOGIES DSOX1202G (serial CN######)
Then after a delay there is a timeout when trying to query ':CHAN3:DISP?'. Since this is a two channel scope, polling channel 3 causes an error. Changing the range(1 to 5) to range(1 to 3) on lines 319 and 326 of oscilliscope.py clears the error but now it is hardcoded for 2 channels.
Since there isn't a standard SCPI command to determine the number of channels on a scope, perhaps channels could be added as a parameter when initializing.
The text was updated successfully, but these errors were encountered:
Hello, I am testing keyoscacquire with a Keysight DSOX1202G and when issuing the
scope = koa.Oscilloscope('LONG ADDRESS')
command, the library responds with:Connected to: KEYSIGHT TECHNOLOGIES DSOX1202G (serial CN######)
Then after a delay there is a timeout when trying to query
':CHAN3:DISP?'
. Since this is a two channel scope, polling channel 3 causes an error. Changing therange(1 to 5)
torange(1 to 3)
on lines 319 and 326 of oscilliscope.py clears the error but now it is hardcoded for 2 channels.Since there isn't a standard SCPI command to determine the number of channels on a scope, perhaps channels could be added as a parameter when initializing.
The text was updated successfully, but these errors were encountered: