Appears to be some more recent instances of mismatched gases in the carbon isotope product. This has been dealt with within NEONiso by including a dataframe allowing for corrections within package internal data.
Suspected mismatched reference/measured values:
BLAN, early 2023, high reference gas
DEJU, late 2022-2023, med reference gas
SOAP, late 2021, all reference gases (might also be a short term issue with measurements, need to investigate more)
SRER, early 2023, low reference gas, CO2 values only
STEI, second half of 2022, high reference gas (change in reference CO2/d13C values but does not appear to be a change in measured values)
Appears to be some more recent instances of mismatched gases in the carbon isotope product. This has been dealt with within NEONiso by including a dataframe allowing for corrections within package internal data.
Suspected mismatched reference/measured values:
(fyi @cflorian1)
The text was updated successfully, but these errors were encountered: